Skip to Content
Capable and Versatile

LS-R740(WL)

The LS-R740(WL) is ideal for testing waterproof components or finished products that require IP protection rating. The part to be tested is placed in a capsule and pressurized to measure the pressure change caused by leakage flowing into the part.

Cosmo LS-R740(WL) Leak Tester for Waterproof Components
Cosmo LS-R740(WL) Intuitive Interface Icon

Intuitive and easy-to-navigate interface with simple menu configuration

Cosmo LS-R740(WL) Waveform Display Icon

Test pressure and differential pressure waveform display

Cosmo LS-R740(WL) Noise Compensation Icon

The influence of the noise generated during the test on the workpiece can be compensated, reducing the test time.

Cosmo LS-R740(WL) Language Selection Icon

Three selectable languages. English, Chinese, Japanese.

Cosmo LS-R740(WL) USB Port Icon

USB port for easy management of test data

Cosmo LS-R740(WL) Sensor Location Icon

Closest location of the sensor unit to the target capsule

Product Features


General Specifications

Differential pressure


Resolution0.1 Pa
Display range± 999.9 Pa
Accuracy± 2.5% of reading ± 1 Pa (500 Pa or less: ± 2 Pa)
Sensor test pressure5 mPa
Unit


Test accuracykPa, MPa (kg/cm2, psi, bar, etc.)
Leakage ratePa, kPa, mL/s, mL/min, L/min, Pa·m3/s, E-3 Pa·m3/s, Pa/s, Pa/min, *Pa/s, *Pa/min (mmH2O, mmHg, inH2O, etc.)
K (Ve) leakage coefficientmL, L (in3, ft3)
Leakage screen4 digits, sampling frequency: 10 times per second.
Leak limit999.99 maximum
Number of channels32 channels (0 to 31)
Timer999.9 seconds (resolution 0.1 seconds)
Power supplyAC100 ~ 240V ± 10%, 50/60Hz, 70VA max.
Ambient temperature5 to 45 ℃
Humidity80% relative humidity or less / no dew condensation
WeightController: approx. 2.6 kg, Pneumatic unit: approx. 7.1 kg
Pressure sourceClean, dry air. The source pressure must be sufficiently higher than the test pressure.
Port sizeThe special gasket is used only for the tubes of the WORK and MASTER ports. Rc (PT) 1/4: pressure source, WORK and MASTER ports.
CPUSH-2A 144 MHz, DRAM 8 MB
RS-232CThe limit values are transmitted, as well as the large and small leak limits. The results of the LL-DET and S-DET tests are transmitted. There are 4 other formats available. Test, leak, test pressure, channel number, timers, etc. Test the parameters of all configuration screens.
USB portM1K1, M3K4, ID2 format, L2 format, others. Test data, test parameters.

Are you looking for more technical information about the LS-R740(WL)?

Download the complete catalog from the manufacturer Cosmo Instruments.